In view of the electromagnetic flow sensor protection grade reaches IP65.
A, measuring, mixing fluid should be installed at the place of fluid unseparated; Measurement of two-component fluid, wait to mix and can measure; When measuring chemical reaction fluid, can wait until fully complete chemical reaction.
Second, installed on the position of the small vibration.
Three, because is the electromagnetic flow sensor, so should avoid electromagnetic interference.
Fourth, avoid negative pressure of tube happen.
Five, the grounding more convenient.
Six, installation environment should not have a high concentration of corrosive gases.
Seven, placed in the shade.
Eight, attention should be paid to the influence of ambient temperature. Average temperature to control within the scope of the 25 to 60 degrees Celsius, some of the more strict flow sensor.
Nine, watch the applicable scope of humidity sensor. Generally in the 10% to 90%.
Ten, prevent the rain.
To leave before eleven, sensor location must be straight.
12, the best, perpendicular to the direction of flow fluid through a sensor from bottom to top.