1 range: the algebraic difference between the upper and lower limits of the measurement range.
	2 accuracy: the measured results are consistent with the true values.
	3 there are usually sensitive components and conversion components.
	A sensing element is a part that can be measured directly (or response) in a sensor.
	The conversion element refers to a sensor that is capable of feeling (or responsive) to the north side of the sensing element and is converted into a portion of the electrical signal that is transmitted and (or) measured.
	When the output is the standard signal, it is called the transmitter.
	4 measurement range: the range of the measured value within the allowable error limit.
	5 repeatability: in all of the following conditions, the degree of conformity between the results obtained by repeated measurements of the measured amount of the same:
	Same measuring method:
	The same observer:
	Same measuring instrument:
	Same place:
	Same conditions of use:
	Repetition in a short period of time.
	6 resolution: sensors in the measurement range of the measurement can be detected by measuring the smallest change.
	7 threshold: the smallest amount of change that can be measured at the output of the sensor.
	8 Zero: the absolute value of the output is the smallest state, such as the equilibrium state.
	9: the linearity of the calibration curve and a limit of consistency.
	The degree of deviation from the calibration curve and the linear deviation of a specified line in the 10 phenanthrene.
	11 long term stability: the sensor can be kept within the specified time is not more than the ability to allow errors.
	12 natural rate: in the absence of resistance, the sensor free (without external force) oscillation with the rate of.
	13 response: the characteristics of the measured changes in output.
	14 the range of compensation temperature: the temperature range of the sensor to keep the range and the zero balance within the prescribed limit.
	15 creep: when the machine is measured to be more than the environmental conditions remain constant, the amount of change in the specified time.
	16 insulation resistance: if no other requirements, refer to the provisions of the DC voltage at room temperature, the provisions of the insulation between the measured value from the sensor.
	17 excitation: the external energy (voltage or current) applied to the sensor to work properly.
	18 maximum excitation: in the city conditions, the maximum value of the excitation voltage or current can be applied to the sensor.
	19 input impedance: short circuit at the output, the input impedance of the sensor.
	20 output: sensor generated and applied to be measured as a function of the relationship between the amount of electricity.
	21 output impedance: in the input side of the short circuit, the output of the sensor measured impedance.
	22 zero point output: the output of the sensor is measured at zero in the city.
	23 lag: in the specified range, when the measured value increases and decreases, the maximum difference occurs in the output.
	24 later: the time delay of the change of the output signal with respect to the change of the input signal.
	25 drift: in a certain time interval, the output of the sensor has finally been measured independent of the amount of change is not required.
	26 zero drift: the variation of the zero output at the specified time interval and indoor conditions.
	27 sensitivity: the ratio of the increment of the output of the sensor and the increment of the corresponding input.
	28 sensitivity shift: the slope of the calibration curve due to the change in sensitivity.
	29 thermal sensitivity drift: sensitivity drift due to changes in sensitivity.
	30 thermal zero drift: zero drift due to temperature variation around.